DHS 157.71(23)(b)4.
4. Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom systems and clock and decayed source activity in the unit's computer.
DHS 157.71(23)(b)5.
5. The name of the person who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check.
DHS 157.71(24)
(24)
Records of periodic spot-checks for gamma stereotactic radiosurgery units. DHS 157.71(24)(b)2.
2. The manufacturer's name, model number and serial number for the gamma stereotactic radiosurgery unit and the instrument used to measure the output of the unit.
DHS 157.71(24)(b)8.
8. Notations indicating the operability of radiation monitors, helmet microswitches, emergency timing circuits, emergency off buttons, electrical interlocks, source exposure indicator lights, viewing and intercom systems, timer termination, treatment table retraction mechanism and stereotactic frames and localizing devices.
DHS 157.71(24)(b)9.
9. The name of the person who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check.
DHS 157.71(25)
(25)
Records of additional technical requirements for mobile remote afterloader units. DHS 157.71(25)(b)2.
2. The manufacturer's name, model number and serial number of the remote afterloader unit.
DHS 157.71(25)(b)3.
3. Notations accounting for all sources before the licensee departs from a facility.
DHS 157.71(25)(b)4.
4. Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom system, applicators and source transfer tubes and source positioning accuracy.
DHS 157.71(26)
(26)
Records of surveys of therapeutic treatment units. DHS 157.71(26)(a)(a) A licensee shall maintain a record of radiation surveys of treatment units made under s.
DHS 157.67 (14) for the duration of use of the unit.
DHS 157.71(26)(b)2.
2. The manufacturer's name, model number and serial number of the treatment unit, source and instrument used to measure radiation levels.
DHS 157.71(26)(b)3.
3. Each dose rate measured around the source while the unit is in the off position and the average of all measurements.
DHS 157.71(27)
(27)
Records of 5-year inspection for teletherapy and gamma stereotactic radiosurgery units. DHS 157.71(27)(a)
(a) A licensee shall maintain a record of the 5-year inspections for teletherapy and gamma stereotactic radiosurgery units required by s.
DHS 157.67 (15) for the duration of use of the unit.
DHS 157.71(27)(b)3.
3. The manufacturer's name and model number and serial number of both the treatment unit and source.
DHS 157.71(28)
(28)
Records of decay of strontium-90 sources for ophthalmic treatments. DHS 157.71(28)(a)
(a) A licensee shall maintain a record of the activity of a strontium-90 source required by s.
DHS 157.65 (6) for the life of the source.
DHS 157.71 History
History: CR 01-108: cr.
Register July 2002 No. 559, eff. — see Note at the start of the chapter;
CR 06-021: r. and recr. (14)
Register October 2006 No. 610, eff. 11-1-06;
CR 16-078: am. (8)
Register January 2018 No. 745, eff. 2-1-18;
CR 22-015: cr. (1) (c), am. (4) (b) 3., (5) (c) Register June 2023 No. 810, eff. 7-1-23. DHS 157.72(1)(a)(a) A licensee shall report to the department any event, except for events that result from intervention by a patient or human research subject, in which the administration of radioactive material or resulting radiation, except for permanent implant brachytherapy, results in any of the following:
DHS 157.72(1)(a)1.
1. A dose that differs from the prescribed dose or dose that would have resulted from the prescribed dosage by more than 0.05 Sv (5 rem) effective dose equivalent, 0.5 Sv (50 rem) to an organ or tissue or 0.5 Sv (50 rem) shallow dose equivalent to the skin and to which any of the following apply:
DHS 157.72(1)(a)1.b.
b. The total dosage delivered differs from the prescribed dosage by 20% or more or falls outside the prescribed dosage range.
DHS 157.72(1)(a)1.c.
c. The fractionated dose delivered differs from the prescribed dose, for a single fraction, by 50% or more.
DHS 157.72(1)(a)2.
2. A dose that exceeds 0.05 Sv (5 rem) effective dose equivalent, 0.5 Sv (50 rem) to an organ or tissue or 0.5 Sv (50 rem) shallow dose equivalent to the skin from any of the following:
DHS 157.72(1)(a)2.b.
b. An administration of a radioactive drug containing radioactive material by the wrong route of administration.
DHS 157.72(1)(a)2.c.
c. An administration of a dose or dosage to the wrong patient or human research subject.
DHS 157.72(1)(a)3.
3. A dose to the skin or an organ or tissue other than the treatment site that exceeds by 0.5 Sv (50 rem) to an organ or tissue and 50% or more of the dose expected from the administration defined in the written directive.
DHS 157.72(1)(am)
(am) For permanent implant brachytherapy, a licensee shall report to the department any event, except for events that result from intervention by a patient or human research subject, in which the administration of byproduct material or radiation from byproduct material, excluding sources that were implanted in the correct site but migrated outside the treatment site, results in any of the following:
DHS 157.72(1)(am)1.
1. The total source strength administered that differs by 20% or more from the total source strength documented in the post-implantation portion of the written directive.
DHS 157.72(1)(am)2.
2. The total source strength administered outside of the treatment site exceeding 20% of the total source strength documented in the post-implantation portion of the written directive.
DHS 157.72(1)(am)5.
5. Sealed source implanted directly into a location discontiguous from the treatment site, as documented in the post-implantation portion of the written directive.
DHS 157.72(1)(am)6.
6. A leaking sealed source resulting in a dose that exceeds 0.5 Sv (50 rem) to an organ or tissue.
DHS 157.72(1)(b)
(b) A licensee shall report to the department any event resulting from intervention of a patient or human research subject in which the administration of radioactive material or radiation therefrom results or will result in an unintended permanent functional damage to an organ or a physiological system, as determined by a physician.
DHS 157.72(1)(c)
(c) A licensee shall notify the department by telephone no later than the next calendar day after discovery of the medical event.
DHS 157.72 Note
Note: Submit report to the Department via telephone at (608) 267-4797 or via facsimile at (608) 267-3695.
DHS 157.72(1)(d)1.1. A licensee shall submit a written report to the department within 15 working days after discovery of the medical event.
DHS 157.72(1)(d)2.g.
g. Whether the licensee notified the person or the person's responsible relative or guardian and if not, why not.
DHS 157.72(1)(d)3.
3. The report required in subd.
1. may not contain the affected individual's name or any other information that could lead to identification of the person.
DHS 157.72 Note
Note: Submit written reports to the Department at: Department of Health Services, Radiation Protection Section, P.O. Box 2659, Madison WI 53701-2659.
DHS 157.72(1)(e)
(e) A licensee shall notify the referring physician of the event and also notify the person who is the subject of the medical event no later than 24 hours after its discovery unless the referring physician personally informs the licensee either that the physician will inform the person or that, based on medical judgement, telling the person would be harmful. A licensee is not required to notify the person without first consulting the referring physician. If the referring physician or the affected person cannot be reached within 24 hours, a licensee shall notify the person as soon as possible thereafter. A licensee may not delay any appropriate medical care for the person, including any necessary remedial care resulting from the medical event, because of any delay in notification. To meet the requirements of this paragraph, the notification of the person who is the subject of the medical event may be made instead to that person's responsible relative or guardian. If a verbal notification is made, a licensee shall inform the person or appropriate responsible relative or guardian that a written description of the event may be obtained from the licensee upon request. A licensee shall provide the written description if requested.
DHS 157.72(1)(f)
(f) If the person who is the subject of the medical event was notified under par.
(d), a licensee shall also furnish within 30 days after discovery of the medical event a written report to the person by sending either of the following:
DHS 157.72(1)(f)2.
2. A brief description of both the event and the consequences as they may affect the person.
DHS 157.72(1)(g)
(g) Aside from the notification requirement, nothing in this subsection affects any rights or duties of a licensee or physician in relation to each other, to any person affected by the medical event or to any individual's responsible relatives or guardians.
DHS 157.72(1)(h)
(h) A licensee shall retain a record of a medical event under s.
DHS 157.71 (4). A copy of the record required under s.
DHS 157.71 (4) shall be provided to the referring physician, if other than the licensee, no later then 15 days after the discovery of the event.
DHS 157.72(2)
(2)
Report of a dose to an embryo or fetus or a nursing child. DHS 157.72(2)(a)(a) A licensee shall report to the department any dose to an embryo or fetus that is greater than 50 mSv (5 rem) dose equivalent that is a result of an administration of radioactive material or radiation from radioactive material to a pregnant individual unless the dose to the embryo or fetus was specifically approved, in advance, by the authorized user.